  1. When one metal is brought in contact with another, a voltage difference occurs due to their differences in electron mobility. When a metal is brought in contact with a solution of salts or acids, a similar electric potential is caused, which has led to the invention of batteries. Similarly, an electric potential develops when one liquid is brought in contact with another one, but a membrane is needed to keep such liquids apart. A pH meter measures essentially the electro-chemical potential between a known liquid inside the glass electrode (membrane) and an unknown liquid outside. Because the thin glass bulb allows mainly the agile and small hydrogen ions to interact with the glass, the glass electrode measures the electro-chemical potential of hydrogen ions or the potential of hydrogen. To complete the electrical circuit, also a reference electrode is needed. Note that the instrument does not measure a current but only an electrical voltage, yet a small leakage of ions from the reference electrode is needed, forming a conducting bridge to the glass electrode. A pH meter must thus not be used in moving liquids of low conductivity (thus measuring inside small containers is preferable) The pH meter measures the electrical potential (follow the drawing clock-wise from the meter) between the  mercuric chloride of the reference electrode and its potassium chloride liquid, the unknown liquid, the solution inside the glass electrode, and the potential between that solution and the silver electrode. But only the potential between the unknown liquid and the solution inside the glass electrode change from sample to sample. So all other potentials can be calibrated out of the equation. The calomel reference electrode consists of a glass tube with a potassium chloride (KCl) electrolyte which is in intimate contact with a mercuric chloride element at the end of a KCL element. It is a fragile construction, joined by a liquid junction tip made of porous ceramic or similar material. This kind of electrode is not easily 'poisoned' by heavy metals and sodium.
    The glass electrode consists of a sturdy glass tube with a thin glass bulb welded to it. Inside is a known solution of potassium chloride (KCl) buffered at a pH of 7.0. A silver electrode with a silver chloride tip makes contact with the inside solution. To minimise electronic interference, the probe is shielded by a foil shield, often found inside the glass electrode.
    Most modern pH meters also have a thermistor temperature probe which allows for automatic temperature correction, since pH varies somewhat with temperature

You're reading: How ph meter works?

Question Stats

Latest activity: earlier.
This question has 1 answers.

BECOME A GUIDE

Share your knowledge and help people by answering questions.
Unanswered Questions